Narashima Reddy B*, Prasad Kulkarni, Nirmala BC and Ravi Shivaraman
Department of Anaesthesiology, MVJ Medical College & research hospital, Hoskote, Bangalore, Karnataka, India.
Received: 03/02/2014 Accepted: 24/02/2014 Revised: 21/02/2014
Visit for more related articles at Research & Reviews: Journal of Medical and Health Sciences
We present a case of a 50 yr old female ASA I, who presented with hemangioma on the right side of the neck was posted for excision. Patient was shifted to operation theatre, intravenous (iv) line secured with 18gauge iv canula. Monitors connected .Induced and intubated. After 30 minutes of induction, we noticed difficulty in ventilation and tight bag since the patient was manually ventilated. There was fall in EtCO2 levels and later saturation. Auscultation revealed silent chest. Since the surgery involved dissection around the dome of right lung, pneumothorax was the first presumptive diagnosis. Patient was immediately ventilated with 100% O2. And a wide bore needle was inserted in right 2nd intercostal space which failed to relieve the resistance to ventilation. Endotracheal tube suctioning was done and the tube cuff was deflated but in vain. Patient was extubated and re-intubated with a fresh tube and the rest of the course of anaesthesia was uneventful. On inspection the extubated tube was found to be full of thick mucous secretions occluding the entire lumen of the tube. We present this case in view of the unusual cause for difficulty in ventilation in an ASA I patient with no history of previous lung infection. In this era of ventilation, incidental use of manual ventilation helped us diagnose the obstruction early. Mechanical obstruction is a rare & unanticipated cause of ‘tight bag’, unless intervened early could be fatal. The surprise is this is a rare cause of blocked ETT, reason of the mucous plug and from where did it come remains enigma
tight bag, ET tube block
Endotracheal tube (ETT) is used to secure airway. When that itself becomes the cause for airway obstruction it becomes difficult to diagnose and act immediatly. Intra operative difficulty in ventilation is a critical incident involving multiple causes leading to confusion and panic when all the causes are ruled out leading to difficulty in treatment. This requires immediate intervention as the situation is fatal.
Here we present a case of a 50 yr old female of 45kg, without any co-morbidity and insignificant past history, presented with hemangioma on the right side of the neck posted for excision. She was accepted under ASA1. Tab alprsolam 0.5mg and tab ranitidine 150mg was given at night before the surgery. Written informed consent was taken for general anaesthesia after explaining the pors and cons.
Patient was shifted to operation theatre, intravenous (iv) line secured with 18gauge canula and started on ringer lactate. Patient was given injection (inj) glycopyrrolate 0.2mg as antisialogogue and Inj morphine 4mg as analgesic 10 minutes before induction. All the monitors were connected – Non Invasive Blood Pressure, pulse oximeter, Electrocardiography, End tidal Carbondioxide (EtCO2) and the basal readings were recorded. Emergency drugs and difficult intubation cart were kept ready. Patient was preoxygenated with 100% O2. Patient was induced with inj. Thiopentone sodium 5mg/kg. Muscle relaxation achieved with inj. Succinyl choline 75mg. She was intubated with endotracheal (ET) tube of 7.5 and secured after confirming bilateral air entry. Anaesthesia was maintained with Oxygen (O2), Nitrous oxide, Halothane and vecuronium. We were manually ventilating the patient with bain’s circuit.
In the middle of the surgery around 30minutes after induction, we noticed difficulty in ventilation and tight bag. There was fall in EtCO2 levels followed by desaturation. Auscultation revealed silent chest bilateraly. Since the surgery involved dissection around the dome of right lung we suspected accidental injury of the pleura leading to pneumothorax. Immediately surgery was stopped and the area of dissection was covered with normal saline, patient was ventilated with 100% O2 & expiratory valve was kept completely open. A wide bore needle attached to a syringe was inserted in right 2nd intercostal space. But this action failed to relieve the resistance for ventilation. Then endotracheal tube suctioning was done and the tube cuff was deflated but in vain. We decided to extubate and mask ventilate as the last resort. We were able to mask ventilate saturation picked up. Patient was re-intubated with a fresh tube and the rest of the course of anaesthesia and surgery was uneventful.
On inspection the extubated tube was found to be full of very thick mucous secretions occluding the entire lumen of the tube at the cuffed area.
Differential diagnosis considered in this case were, pneumothorax, bronchospasm, ET tube kinking, bevel touching carina, herniation of endotracheal tube cuff, and foreign body obstruction of the tube.
A complete obstruction of the airway in an intubated patient is quite rare. Usually we see partial and gradual obstruction, with high inspiratory pressure during controlled ventilation. In these cases there will be gradaual fall in EtCO2, but saturation will be maintained. Possible causes for this will be poor lung compliance, acute bronchospasm, obstuction or kinking [1] of the tube and circuit, external compression,tention pneumothorax, endobronchial mass, foreign body or thick secretions in the ET tube, biting, beavel abetting the tracheal wall, obstruction by cuff system malfunction.
Several cases have been reported with cuff [2] defects leading to herniation and intraluminal obstruction. It is possible for an ET tube cuff, if overinflated or defective, to herniate over the end of the tube resulting in partial or total airway obstruction. If this happens the airway should be replaced as a cuff deformed in this way may herniate again.
There are several reported cases on airway obstruction due to blood clots by Arney et al [3]. Hyun Kyoung Lim et.al [4] experienced a case of complete obstruction of the endotracheal tube due to a blood clot during surgery with a patient in the prone position who had a hemo-pneumothorax. But there are few reported cases with mucus plug obstruction.
Endotracheal tubes can become obstructed with mucus, particularly in cases where the mucus contains blood, secretions are thick and when there is inadequate humidification [5]. It is possible to have an endotracheal tube totally obstructed by mucus and the patient cannot breath yet it will be able to pass a suction catheter through the viscid obstruction. This had occured in over case. If there is any doubt of airway patency then the airway should be replaced immediately. But patients should never be extubated unless some help is present who can assist in immediate reinsertion of an endotracheal tube, perhaps of a smaller diameter than the one removed.
Whenever an endotracheal tube obstruction is considered as the possibility, we have to evaluate the tracheobronchial tree with a flexible bronchoscope for direct detection of bleeding sites and for the removal of any visible blood clots or mucus, even in cases where there is no evidence of preoperative hemoptysis or infection [6]. Additionally, it should be kept in mind that a blood clot or mucus plug can migrate into the endotracheal tube due to intermittent positive pressure ventilation. We also should keep in mind that even though endotracheal suction applied may not be able to eliminate secretions [7-10].
We present this case in view of the unusual cause for difficulty in ventilation in an ASA I patient with no history of previous lung infection. In this era of ventilation with ventilators, use of manual ventilation helped us to diagnose the obstruction early. Mechanical obstruction is a rare & unanticipated cause of ‘tight bag’, which unless intervened early could be fatal. Failure to detect any cause compelled us to change the tube and after which the course of anaesthesia was uneventful.