ISSN ONLINE(2319-8753)PRINT(2347-6710)

All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Research Article Open Access

Development and Validation of LC-MS/MS Method for the Analysis of P-Toluenesulfonicacid, In Antiviral Drug Acyclovir

Abstract

A sensitive and selective liquid chromatographic-tandem mass spectrometric (LC/MS/MS) method was developed and validated for the trace analysis (>1 ppm level) of p-Toluenesulfonicacid, a genotoxic impurity, in acyclovir drug substances. The chromatographic separation was achieved on Symmetry C-18 (150 X 4.6mm, 3.5μm) column using a mobile phase consisting of 5 mM ammonium acetate and methanol (70:30 v/v) at flow rate of 0.4 mL /min. The API-4000 LC/MS/MS was operated on an electrospray in negative mode. The selected ion monitoring (SIM) mode was used during the analytical run and it is able to quantitate up to 1.5 ppm of p-Toluenesulfonicacid. The newly developed method was validated as per achieved on International Conference on Harmonization (ICH) guidelines

Noor Shabana, K Dayananda Reddy

To read the full article Download Full Article | Visit Full Article